If you have any question, please contact us
Introduction:
Pneumatically driven and electrically controlled system
Built-in ventilator, 5.4’’ LCD display screen
Closed; Semi-Closed; Semi-Open working mode
Applied for Adult and Pediatric
Sterile, abrasion resistance to eliminate bacterial penetration
Transport moisture from the skin to the surface of the fabric, keeping the dresser cool, dry and comfortable.
Breathable, light weight and soft feeling
Operating Modes: Pneumatically driven & electronically controlled, closed, semi-closed and semi-open
Anesthesia Ventilator: Built-in ventilator, 5.4" LCD display screen
Standard integrated metallic modular respiratory circuit; completely solves the problem of gas sealing under
high-temperature disinfection, easy to clean and disinfect
Equipped with electronic pressure differential flow sensor, imported respiratory control valve
Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from O2 & NO2: 0.32~0.6 MPa.
4-tubes Flowmeters, O2 & N2O; 0.1L/Min ~ 10L / Min; Oxygen Flush: 25~75 L / Min
Respiration Mode: A/C,IPPV,SIPPV,IMV,SIMV,MANUAL
Tidal Volume: 0~1500ml
I/E Ratio: 4:1 ~ 1:10
One vaporizer: Enflurane/ Isoflurane/Sevoflurane/Halothane
Vaporizer Setting Range: 0.5%~5%
Respiration Frequency : 1~99 bpm;
Ptr: -1.0kPa ~ 1.0 kPa
CO2 absorber capacity : 1 kg
Optional: Patient monitor
Alarm parameters: upper/lower limit of airway pressure,tital volume exceed limitation,no tidal volume output,apnea,AC power malfunction,Battery low power.
Vaporizer: Imported tank components assembled locally; Enflurane, Isoflurane and Sevoflurane (Halothane optional). Also possesses self-compensating functionality
Multi-parameter Patient Monitor: Optional
High precision four-tube flowmeter, fitted with interlocking oxygen-nitrous oxide supply, guarantees oxygen output is never lower than 21%
Nonhazardous environmentally-friendly frame, with automated self-checking function during startup and an automatic self-calibrating sensor
Power Failure Support Functions: Automatically converts to backup power mode during AC power outages
Alarms: Audible and/or visual alarms for tidal volume, airway pressure, asphyxia, power failure, oxygen failure, and other alarms, also features an alarm query function
Equipped with an advanced international rotating Na-lime tank, uses imported PPSU material and is temperature resistant to 134°C
Main technical specifications: | |
Gas requirement | Medical oxygen and nitrous oxide with a pressure ranging from O2 & NO2: 0.32~0.6 MPa. |
4-tubes Flowmeters, O2 & N2O | 0.1L/Min ~ 10L / Min |
Oxygen Flush | 25~75 L / Min |
Respiration Mode | A/C,IPPV,SIPPV,IMV,SIMV,MANUAL |
Tidal Volume | 0~1500ml |
I/E Ratio: | 4:1 ~ 1:10 |
One vaporizer | Enflurane/ Isoflurane/Sevoflurane/Halothane |
Vaporizer Setting Range | 0.5%~5% |
Respiration Frequency | 1~99 bpm |
Ptr | -1.0kPa ~ 1.0 kPa |
CO2 absorber capacity | 1 kg |
Optional | Patient monitor |
Alarm parameters | upper/lower limit of airway pressure, tital volume exceed limitation,no tidal volume output,apnea, AC power malfunction,Battery low power. |