5.4” LCD Display Screen
4 Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min
Vaporizer 2 Vaporizers,(Enflurane/Isoflurane/Sevoflurane, Option: Halothane)
Respriation Mode:A/C, IPPV, SIPPV, IMV, SIMV, MANUAL
Tidal volume 0-1500ml
Respriation Frequency 1-99bmp
I/E Ratio :4:1 ~ 1:10, Has the inverse ratio ventilation
Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from O2 & NO2: 0.32~0.6 MPa.
4-tubes Flowmeters, O2 & N2O; 0.1L/Min ~ 10L / Min; Oxygen Flush: 25~75 L / Min
Respiration Mode: A/C,IPPV,SIPPV,IMV,SIMV,MANUAL
Tidal Volume: 0~1500ml
I/E Ratio: 4:1 ~ 1:10
One vaporizer: Enflurane/ Isoflurane/Sevoflurane/Halothane
Vaporizer Setting Range: 0.5%~5%
Respiration Frequency : 1~99 bpm;
Ptr: -1.0kPa ~ 1.0 kPa
CO2 absorber capacity : 1 kg
Optional: Patient monitor
Alarm parameters: upper/lower limit of airway pressure,tital volume exceed limitation,no tidal volume output,apnea,AC power malfunction,Battery low power.
Physical specifications
|
|
Screen
|
5.4'' Display Screen
|
Suitable
|
Adult & Child
|
Mode
|
pneumatically powered and electrically controlled system
|
Working Mode
|
Closed; Semi-Closed; Semi-Open
|
Circuit
|
Breathing circuit integrated standards
|
Tube
|
4Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min
|
Trolley
|
Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated
brake provisions |
Drawer unit
|
ONE drawer totally extractable
|
Technical Specifications
|
|
Gas requirement
|
Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa.
|
safety valve
|
<12.5 kPa
|
Respiratory Rate
|
1~99bpm
|
Flow meter
|
O2: 0.1~10L/Min, N2O: 0.1~10L/Min
|
oxygen concentration in the mixed gas N2O/O2
|
> 21%
|
Oxygen Flush
|
25~75 L/min
|
modes of ventilation
|
A/C , IPPV, SIPPV, IMV, SIMV, MANUAL
|
I/E Ratio
|
4:1 ~ 1:10, Has the inverse ratio ventilation
|
Tidal Volume
|
0~1500ML
|
ptr inspiratory triggering pressure
|
-0.1kPa~1.0kPa
|
Inspiratory Plateau
|
0~1s
|
Maximum safety pressure
|
≤ 12.5 kPa
|
Pressure limit range
|
0 ~ 6.0 kPa
|
Airway pressure alarm: Audible and visual and with yellow and red color indicating
|
Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
|
±0.2 kPa
|
|
Tidal volume alarm: Audible and visual and with yellow and red color indicating
|
upper alarm: 50 to 2000ml, lower alarm: 0~1800ml
|
Power Supply Alarm
|
Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s |
The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
|
|
Vaporizer
|
|
Vaporizer
|
Enflurane/ Isoflurane(Option: Halothane)
|
Setting Range: 0.5~5%
|
|
Safty & Alarm
|
|
Oxygen Alarm
|
It alarms when the oxygen supply from pipe or cylinders lower than 0.2MPa
|
Ventilation Volume Alarm
|
Lower: 0~12L/Min; Up: 18L/ Min
|
Power Alarm
|
It Alrmas while AC and DC supply failure ; Keep alraming time: >120s
|
Air Tract Pressure Alarm
|
Lower: 0.2kPa ~5.0 kPa; Up: 0.3kPa ~ 6.0kPa
|